Instrument Features
1: Instrument minimum detection limit: 0.05% of full-scale signal value
2: Simultaneous detection of gold conjugate fluorescence, intelligently recognizing gold conjugate fluorescent reagent cards
3: Instrument stability: 10 measurements, TC ratio CV less than 0.5%
4: Instrument signal-to-noise ratio: 0.01% (1:10000)
5: Test speed: maximum 120T/hour (external incubation)
6: Multiple projects can be tested simultaneously, with temperature control in the built-in incubation zone, eliminating the need for manual timing and waiting
7: Automatic identification of items, automatic incubation timing, automatic detection, automatic card ejection, and automatic result uploading
The 8:ld chip reads project parameters and detects the barcode of the cartridge
9: Support the automatic generation of Excel files for result export
10: Supports network connections, WIFI and 4G
(1. WIFI connection: Select the WIFI hotspot on the LCD screen and enter the password
2. 4G connection: Configure a data card to automatically connect to the base station
3. Automatic fault alarm and remote diagnosis
4. Automatic download of standard curve
5. Automatic program updates and remote fault diagnosis
Basic Parameters
Dimensions (LxWxH): 320*315*300(mm), customizable
Weight: ≤7.5kg
True color touch LCD screen 7-inch capacitive touch
LCD resolution: 800x480 pixels
Power supply: AC220V, 50Hz, 24V/4A
Instrument interface: USB, RS-232, LAN, connectable to LIS
Storage results: 8000 entries
Operating mode: Stand-alone or connected to PC software
Project identification: Intelligent identification of 1200 types of projects and batches
Number of detection channels: 6 channels
Number of test lines for a single channel: Maximum support for 5 lines
Quality Control Calibration Internal Quality Control Calibration
Fluorescence excitation/emission wavelength band: 365/610(nm) (customizable)
Colloidal gold wavelength band: 525(nm)